Common Pitfalls in Nanoparticle Analysis

Modern nanoparticle analyzers are designed for ease of use, delivering high accuracy for idealized, uniform samples.

However, their streamlined workflows rely on baseline assumptions that
often fail when applied to complex, real-world nanoparticles.
Here's how you can unlock the full potential of nanoparticle analyzers:

01

Establish Your Instrument's True Limitations

The single most important step is to establish the actual performance limitations of your instrument for your specific sample. Instrument manufacturers often provide specifications, but these claims are frequently based on ideal standards like polystyrene "beads".

 

The reality is that your sample's unique complexity—be it its composition, morphology (hollow vs. solid), or fluorescence—will define your true detection limit. For example, a scattering-based detection limit advertised as 50 nm for solid beads may in fact be 100 nm or higher for hollow nanoparticles like vesicles due to their lower refractive index.​

02

Understand the Role of Standards in Validation

​The best approach to understanding your instrument's real-world capabilities is to validate it. Always begin by setting up the instrument according to the manufacturer's recommendations to ensure baseline QC passes.

 

The next step, often disregarded, is to run a known reference material. Ideally, this standard should have properties as close as possible to your actual sample. Understandably, having a perfect in-house standard isn't always feasible, especially during early-stage R&D, therefore using well-characterized commercial standards like Syncles™ is a critical alternative.

03

Standardize for Meaningful Data Comparison

In multi-user facilities or collaborative projects, data variability is the is the primary barrier to precision. Without standardized protocols, results become operator-dependent, influenced by subjective settings like camera/shutter settings, detector thresholding, or manual gating.

 

To ensure data integrity, every operator should calibrate with the same invariant reference material prior to measurement. This simple step creates a universal baseline, transforming isolated datasets into robust evidence that stands up to scrutiny across different labs and timepoints.

How Standards Overcome Pitfalls

PITFALL 01
The "Invisible" Nanoparticles

Small particles often go undetected below an instrument's Limit of Detection (LOD). This causes the instrument to report an artificially narrow size distribution with shifted peak, as if the smaller particles didn't exist in the sample!

“True” size distribution Instrument limitation “Measured” size distribution Size Size detection limit Size Size (i) (ii) (iii) (iv)
SOLUTION 01
Establish True LOD

By running a series of size standards that resemble your sample, you can empirically find the exact "cutoff" point. This establishes the true resolution limit for your specific material.

Limit of detection (LOD) established to be ~60 nm Size Size (i) (ii)
PITFALL 02
The "Merged" Peaks

An analyzer may not resolve two distinct but closely-sized populations (e.g. 80 nm and 100 nm). It incorrectly reports a single broad blob instead of two distinct peaks.

Size Instrument limitation “Measured” size distribution Size Size “True” size distribution (i) (ii) (iii) (iv)
SOLUTION 02
Define Resolution Limit

By running a series of size standards that resemble your sample, you can empirically find the exact "cutoff" point. This establishes the true resolution limit for your specific sample.

Size Size
PITFALL 03
Poor Calibration

A poorly calibrated detector fails to distinguish dim nanoparticles from background noise. The readout only captures the brightest outliers, misrepresenting your sample.

Fluorescence Fluorescence Fluorescence (i) (ii) (iii) (iv)
SOLUTION 03
Optimize Signal-to-Noise

By running fluorescent/brightness standards similar to your samples, you can confidently adjust detector gain to lift your true population above the noise floor.
 

Blue fluorescence Blue fluorescence Blue fluorescence Green fluorescence Yellow fluorescence Orange fluorescence Red fluorescence Instrument noise Poor SNR Good SNR Repeat above for all other fluorescent detectorsusing different fluorescent standards (i) (ii)

The Goal:
Clarification
and Confidence

Without proper controls, non-expert users risk mistaking flawed output for reliable data.​

Verify accuracy to ensure data integrity

Clarify actual limits for your specific sample.

Know exactly when you can and cannot trust your results.